Standard ventilation psychrometer is designed according to the principle of ventilation psychrometer. It is equipped with two temperature sensors, one for dry bulb temperature measurement and the other for wet bulb temperature measurement. The wet bulb temperature sensor is wrapped with gauze soaked in water. The evaporation amount of the wet bulb surface of the humidity temperature sensor depends on the relative humidity of the atmosphere, and the heat required for evaporation causes a decrease in the temperature of the wet bulb. The relative humidity is calculated by the humidity formula based on the factors, such as dry bulb temperature, wet bulb temperature and atmospheric pressure at that time. The motor drives the fan to ventilate the surface of the dry and wet bulb temperature sensor, so that the surface of the sensor can maintain a substantially constant wind speed. Thus, the humidity measurement will be more stable and accurate.

Key words:

Product Description

Product Overview

Standard ventilation psychrometer is designed according to the principle of ventilation psychrometer. It is equipped with two temperature sensors, one for dry bulb temperature measurement and the other for wet bulb temperature measurement. The wet bulb temperature sensor is wrapped with gauze soaked in water. The evaporation amount of the wet bulb surface of the humidity temperature sensor depends on the relative humidity of the atmosphere, and the heat required for evaporation causes a decrease in the temperature of the wet bulb. The relative humidity is calculated by the humidity formula based on the factors, such as dry bulb temperature, wet bulb temperature and atmospheric pressure at that time. The motor drives the fan to ventilate the surface of the dry and wet bulb temperature sensor, so that the surface of the sensor can maintain a substantially constant wind speed. Thus, the humidity measurement will be more stable and accurate. 

 

Model

NM-HUMI700

Temperature Range

0℃~ 100℃

Humidity Range

10%RH ~ 99%RH

Temperature Resolution 

0.01℃

Humidity Resolution

0.01%RH

Temperature Accuracy

≤ ±0.08℃

Humidity Accuracy

10%RH ~ 70%RH:±1.0%RH;大于 70%RH:±1.7%RH(20℃)

Temperature Transducer Precision platinum resistance thermometer
Display LCD 
Precision Grade Grade II humidity standard 
Communication Interface RS232 interface (optional) 
Applicable Standards

Verification Regulation of Mechanical Thermo - Hygrometers (JJG 205-2005) 
Verification Regulation of Electric Ventilation Psychrometers (JJG 993-2018)

 

  • 产品描述
    • Commodity name: NM-HUMI700 Standard ventilation psychrometer
    • Commodity ID: 1327233533134655488

    Standard ventilation psychrometer is designed according to the principle of ventilation psychrometer. It is equipped with two temperature sensors, one for dry bulb temperature measurement and the other for wet bulb temperature measurement. The wet bulb temperature sensor is wrapped with gauze soaked in water. The evaporation amount of the wet bulb surface of the humidity temperature sensor depends on the relative humidity of the atmosphere, and the heat required for evaporation causes a decrease in the temperature of the wet bulb. The relative humidity is calculated by the humidity formula based on the factors, such as dry bulb temperature, wet bulb temperature and atmospheric pressure at that time. The motor drives the fan to ventilate the surface of the dry and wet bulb temperature sensor, so that the surface of the sensor can maintain a substantially constant wind speed. Thus, the humidity measurement will be more stable and accurate.

    Product Overview

    Standard ventilation psychrometer is designed according to the principle of ventilation psychrometer. It is equipped with two temperature sensors, one for dry bulb temperature measurement and the other for wet bulb temperature measurement. The wet bulb temperature sensor is wrapped with gauze soaked in water. The evaporation amount of the wet bulb surface of the humidity temperature sensor depends on the relative humidity of the atmosphere, and the heat required for evaporation causes a decrease in the temperature of the wet bulb. The relative humidity is calculated by the humidity formula based on the factors, such as dry bulb temperature, wet bulb temperature and atmospheric pressure at that time. The motor drives the fan to ventilate the surface of the dry and wet bulb temperature sensor, so that the surface of the sensor can maintain a substantially constant wind speed. Thus, the humidity measurement will be more stable and accurate. 

     

    Model

    NM-HUMI700

    Temperature Range

    0℃~ 100℃

    Humidity Range

    10%RH ~ 99%RH

    Temperature Resolution 

    0.01℃

    Humidity Resolution

    0.01%RH

    Temperature Accuracy

    ≤ ±0.08℃

    Humidity Accuracy

    10%RH ~ 70%RH:±1.0%RH;大于 70%RH:±1.7%RH(20℃)

    Temperature Transducer Precision platinum resistance thermometer
    Display LCD 
    Precision Grade Grade II humidity standard 
    Communication Interface RS232 interface (optional) 
    Applicable Standards

    Verification Regulation of Mechanical Thermo - Hygrometers (JJG 205-2005) 
    Verification Regulation of Electric Ventilation Psychrometers (JJG 993-2018)

     

    Key words: